    Eppendorf PICASO Pipette Calibration Software


    Eppendorf PICASO Pipette Calibration SoftwareLaboratories in the diagnostic and research sectors demand the utmost accuracy and precision from liquid handling devices. GLP guidelines stipulate that error limits should be checked at regular intervals.

    Pipettes can be used over a long period of time without requiring maintenance; however, leaks may occur as a result of decomposition of the piston grease caused by frequent pipetting of corrosive liquids. It is essential to recognize this problem before this decomposition occurs. The Eppendorf PICASO Pipette Calibration Software monitors and documents the accuracy and precision of a pipette throughout its entire operating life. This new program was designed for recording and processing measuring data from gravimetric or photometric pipette tests. It can be used to test all pipette types from various manufacturers and other liquid handling devices such as a bottletop dispenser.

    Application

    • Software for calibrating piston-stroke pipettes

    Product features

    • User-friendly functions under Microsoft® Windows®
    • Detailed online operating manual and help functions
    • All measurements stored completely, including all individual values
    • No restriction on the number of pipettes, time, and other factors
    • Data transferable to other Microsoft Windows programs (e.g. Microsoft Excel)
    • Can be connected to different balances, or to freely definable balance interface
    • Measurement carried out in line with all relevant guidelines (e.g. DIN, ISO, ASTM)

    Calibration display

    • Choice of pipette type and volumes (via quick-sort function for filters)
    • Choice of guidelines for measurement (test status)
    • Ambient conditions can be entered
    • Pre-measurement possible via icon
    • User guided through measurements
    • Test protocol printout

    Balances

    • Sartorius® RC, MC, AC, BP
    • Mettler® AT
    • Ohaus® AP
    • AnD HM
    • Note: balances not listed here can be adapted by adjusting data transmission accordingly

    Minimum system requirements

    • Computer
    • Intel-Pentium /100 Hz – 32 MB RAM – Windows 95/98/NT/2000
    • CD-ROM drive
    • Two- or three-key mouse
    • RS-232 interface
    • Interface for printer