Eppendorf Refrigerated Multipurpose Centrifuge Model 5810R |
|
Rotors and Adaptors for Multipurpose Centrifuges 5804, 5804 R, 5810, and 5810 R
Eppendorf Multipurpose Centrifuges accommodate fixed-angle and swing-bucket rotors for a variety of applications. A large selection of Adaptors allows maximum application flexibility with a minimum number of rotors.
Product features
- All rotors, buckets, and Adaptors are autoclavable at 121°C
- Stackable modular Adaptors for swing-bucket rotors are easy to disassemble for vessels of different lengths, for cleaning and autoclavin
g
- Adaptors are color coded for quick and simple identification
- Buckets can be easily sealed with aerosol-tight covers to provide protection against biohazardous samples
|
Quick Find links:
|
|
for 5810 and 5810R Only
|
High-Speed Fixed-angle 30-place Microcentrifuge Rotors
- Spin up to 20,800 x g
- Quick release lid

Aerosol-tight Fixed-angle Rotor, for microcentrifuge tubes
Type FA-45-30-11

Fixed-angle Rotor,
for microcentrifuge tubes Type F-45-30-11 |
Specifications |
Catalog No. – Standard – Aerosol-tight |
022637002 022637100 |
| Maximum Capacity |
30 x 2.0 ml |
| Maximum Speed |
14,000 rpm |
| Maximum RCF |
20,800 x g |
| Maximum Radius |
95 mm |
| Maximum Vessel Diameter |
11.5 mm |
| Maximum Vessel Length |
50 mm |
- Max. rotational speed: 14,000 rpm
- Max. rotational speed for 4°C (5804 R/5810 R): 14,000 rpm
- Temperature at max. rotational speed (5804 R, 230 V/5810 R, 230 V): <4°C*
| * |
23°C room temperature and 1 hour run time.
|
 |
Adaptors, aerosol-tight covers, and rotors are autoclavable: 20 min. at 121°C. |
High-speed Fixed-angle Rotor for 6 x 85 ml Tubes
- Adaptors accommodate 1.5 to 50 ml tubes
- Spins conical tubes at high speeds

High-speed Fixed-angle Rotor, 6 x 85 ml
Type F-34-6-38 |
Specifications |
| Catalog No. |
022637207 |
| Maximum Capacity |
6 x 85 ml |
| Maximum Speed |
11,000 rpm* |
| Maximum RCF |
15,500 x g* |
| Maximum Radius |
115 mm |
| Maximum Vessel Diameter |
38 mm |
| Maximum Vessel Length |
120 mm |
| *12,000 rpm/18,500 x g in Model 5810 R. To maintain 4°C with this rotor, a maximum speed of 10,500 rpm is required. |
| Tube |
|
(Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.) |
 |
 |
 |
 |
| Tube capacity (ml) |
1.5–2 |
7–15 |
15 ml conical |
15–18 |
| Max. tube length for all positions below the rotor lid (mm)* |
110 |
112 |
123 |
123 |
| Adaptor for 6 ml x 85 ml fixed-angle rotor for use with Standard, conical and micro test tubes |
| Adaptor |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
|
| Catalog No. |
022637215 |
022637223 |
022637274 |
022637231 |
| Number
of tubes per Adaptor/per rotor |
4/24 |
2/12 |
1/6 |
1/6 |
| Adaptor bore (dia. mm) |
11 |
16 |
17 |
18 |
| Max.
RCF (g-force) at 11,000 rpm |
15,290 |
15,150 |
14,470 |
14,750 |
| Max.
RCF (g-force) at 12,000 rpm – 5810 R |
18,190 |
18,030 |
17,230 |
17,550 |
| Centrifugation radius (mm) |
113 |
112 |
107 |
109 |
| Tube |
| (Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.) |
 |
 |
 |
 |
| Tube capacity (ml) |
20–30 |
50 |
50 ml conical |
85 |
| Max. tube length for all positions below the rotor lid (mm)* |
123 |
123 |
121 |
121 |
| Adaptor for 6 ml x 85 ml fixed-angle rotor for use with Standard, conical and micro test tubes |
| Adaptor |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Can be inserted without Adaptor |
| Catalog No. |
022637240 |
022637258 |
022637266 |
— |
| Number of tubes per Adaptor/per rotor |
1/6 |
1/6 |
1/6 |
–/6 |
| Adaptor bore (dia. mm) |
26 |
29 |
29.5 |
38 |
| Max. RCF (g-force) at 11,000 rpm |
14,880 |
15,150 |
14,610 |
15,560 |
| Max. RCF (g-force) at 12,000 rpm – 5810 R |
17,710 |
18,030 |
17,390 |
18,500 |
| Centrifugation radius (mm) |
110 |
112 |
108 |
115 |
- Temperature at max. rotational speed (12,000 rpm), 5810 R (230 V): <10°C**
- Temperature at max. rotational speed (12,000 rpm), 5810 R (120 V): <12°C**
- Max. rotational speed for 4°C, 5810 R: 10,500 rpm
- Temperature at max. rotational speed (11,000 rpm), 5804 R: <6°C**
- Max. rotational speed for 4°C, 5804 R: 10,500 rpm
| * |
Exceeding the maximum tube length may reduce the loading capacity of the Adaptor. |
| ** |
23°C room temperature and 1 hour run time. |
 |
 |
| |
Adaptors, aerosol-tight covers, and rotors are autoclavable 20 min. at 121°C. |
Horizontal Drum Rotor
- Accepts up to 60 x 1.5 ml or 120 x 0.4 ml tubes for high throughput
- Carriers can be used as tube racks for easy handling

Drum Rotor,
Type T-60-11 |
Specifications |
| Catalog No. |
022638505 |
| Maximum Capacity |
60 x 1.5–2.0 ml, 120 x 0.4 ml |
| Maximum Speed |
14,000 rpm |
| Maximum RCF |
16,435 x g |
| Maximum Radius |
750 mm |
| Maximum Vessel Diameter |
11 mm |
| Maximum Vessel Length |
41 mm |
PCR/MTP/DWP Swing-bucket Rotor with Two Buckets
- Ideal for deepwell blocks and purification kit assemblies with heights greater than 45mm
- Use with Eppendorf Perfectprep® Plasmid 96 Spin Kits
- Spins up to 10 microplates or 4 deepwell plates

Deepwell plate Rotor, includes
2 Deepwell plate buckets
Type A-2-DWP |
Specifications |
| Catalog No. |
022638564 |
| Maximum Capacity |
2 Deepwell Assemblies (filter and catch-plate stacks) |
| Maximum Speed |
3,700 rpm |
| Maximum RCF |
2,250 x g |
| Maximum Radius |
147 mm |
| Maximum Vessel Dimensions, W x D |
91 x 134 mm |
| Maximum Height |
85 mm |
| Type |
Microplate |
Cell culture |
DWP |
Kits |
 |
 |
 |
 |
 |
| Number per bucket/per rotor |
5/10 |
4/8 |
2/4 |
1/4 |
| Max. RCF (g-force) |
2,250 |
2,250 |
2,250 |
2,250 |
| Max. Centrifugation radius (mm) |
147 |
147 |
147 |
147 |
| Dimensions: Length x Width x Height (mm) |
128 x 85.6 x 15 |
128 x 85.6 x 21 |
128 x 85.6 x 45 |
128 x 85.6 x 60 |
- Max. rotational speed: 3,700 rpm
- Max. rotational speed for 4°C (5804 R / 5810 R): 3,700 rpm
- Temperature at max. rotational speed (5804 R / 5810 R): <0°C**
| * |
The deepwell plate rotor 22 63 856-4 is delivered with buckets of one weight category. When placing follow-up orders for a deepwell plate bucket, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In the case of plates used in deepwell plate buckets for rotor 22 63 856-4 (commercially available microplates, culture plates, and deepwell plates), the maximum weight must not exceed 380 g and the maximum height must not exceed 85 mm. If plates other than the above-mentioned plates are used, the user must consult Eppendorf to confirm the suitability of using these plates. |
| ** |
23°C room temperature and 1 hour run time. |
| |
|
| |
Buckets and rotors are autoclavable 20 min. at 121°C. |
|
Swing-bucket Rotor with Four 100 ml Buckets
- Adaptors accommodate 0.25 to 100 ml tubes
- Accepts four 2 x 50 ml buckets to spin eight 50 ml conicals>
- Spins up to 4,500 x g for quick separation

Swing-bucket Rotor,for
standard tubes, includes 4 x 100 ml rectangular buckets
Type A-4-44 |
Specifications |
| Catalog No. |
022637401 |
| Maximum Capacity |
4 x 100 ml |
| Maximum Speed |
5,000 rpm |
| Maximum RCF |
4,500 x g |
| Maximum Radius |
161 mm |
| Maximum Vessel Diameter |
46 mm |
| Maximum Vessel Length |
125 mm |
| Tube |
|
|
 |
 |
 |
 |
 |
| Tube capacity (ml) |
3–5 |
1.5–2 |
4–7 |
3–15 |
7–17 |
| Typical tube size** |
10 x 75 |
11 x 38 |
12 x 75 or 13 x 100 |
16 x 100 |
17 x 100 |
| Max. tube length with aerosol-tight cover*** |
102 |
105 |
106 |
106 |
106 |
| Max. tube length without aerosol-tight cover*** |
105 |
111 |
108 |
108 |
110 |
| Adaptor for 100 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
|
Set of 2
|
Set of 2
|
Set of 2
|
Set of 2
 |
| Catalog No. |
022637509 |
022637525 |
022637541 |
022637568 |
022637584 |
| Number of tubes per Adaptor/per rotor |
14/56 |
12/48 |
9/36 |
7/28 |
6/24 |
| Adaptor bore (dia. mm) |
11 |
11 |
13 |
16 |
17.5 |
| Max. RCF (g-force) |
4,192 |
4,136 |
4,192 |
4,192 |
4,192 |
| Centrifugation radius (mm) |
150 |
148 |
150 |
150 |
150 |
| Tube |
 |
 |
For use with aerosol-tight cover
 |
|
 |
 |
| Tube capacity (ml) |
15 ml conical |
15 ml conical |
7–18 |
12–30 |
50 ml conical |
| Typical tube size** |
17 x 120 |
17 x 120 |
18 x 100 |
25 x 92 |
30 x 115 |
| Max. tube length with aerosol-tight cover*** |
109 |
121 |
104 |
100 |
112 |
| Max. tube length without aerosol-tight cover*** |
121 |
121 |
107 |
108 |
122 |
| Adaptor for 100 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
|
Set of 2
|
Set of 2
 |
Set of 2
|
Set of 2
|
| Catalog No. |
022637606 |
022637614 |
022637622 |
022637649 |
022637665 |
| Number of tubes per Adaptor/per rotor |
4/16 |
2/8 |
4/16 |
2/8 |
1/4 |
| Adaptor bore (dia. mm) |
17.5 |
17.5 |
20 |
26 |
31 |
| Max. RCF (g-force) |
4,332 |
4,388 |
4,192 |
4,192 |
4,332 |
| Centrifugation radius (mm) |
155 |
157 |
150 |
150 |
155 |
| Tube |
 |
For use with aerosol-tight cover
|
 |
 |
 |
|
| Tube capacity (ml) |
50 ml conical |
50 ml conical |
30–50 |
50–75 |
80–120 |
| Typical tube size** |
30 x 115 |
30 x 115 |
30 x 107 |
34 x 100 |
44 x 100 |
| Max. tube length with aerosol-tight cover*** |
119 |
— |
108 |
108 |
100 |
| Max. tube length without aerosol-tight cover*** |
119 |
119 |
122 |
119 |
114 |
| Adaptor for 100 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
 |
Bucket
|
Set of 2
|
Set of 2
|
Set of 2
 |
| Catalog No. |
022637673 |
022637452 |
022637681 |
022637703 |
022637720 |
| Number of tubes per Adaptor/per rotor |
1/4 |
2/8 |
1/4 |
1/4 |
1/4 |
| Adaptor bore (dia. mm) |
31 |
— |
31 |
35 |
45 |
| Max. RCF (g-force) |
4,388 |
4,500 |
4,192 |
4,192 |
4,192 |
| Centrifugation radius (mm) |
157 |
161 |
150 |
150 |
150 |
- Max. rotational speed: 5,000 rpm
- Max. rotational speed for 4°C (5804 R / 5810 R): 4,800 rpm
- Temperature at max. rotational speed (5804 R / 5810 R): 6°C****
| * |
Swing-bucket rotors are delivered with rectangular buckets of one weight category. When placing follow-up orders for rectangular buckets, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In a rectangular bucket, the overall weight of the Adaptors and test tubes used must not exceed 310 g. |
| ** |
Diameter x length in mm. |
| *** |
Exceeding the maximum tube length may reduce the loading capacity of the Adaptor. Please carry out a manual swing test. |
| **** |
23°C room temperature and 1 hour run time.
|
| |
Adaptors, aerosol-tight covers, and rotors are autoclavable: 20 min. at 121°C. |
PCR Strip Rotor
- Max. rotational speed 10,000 rpm
- Up to 6 x 8-tube strips, 6 x 5-tube strips or 48 x 0.2 ml PCR single tubes
- Safe and easy handling without additional lid
 Type F-45-48-PCR |
Specifications |
| Catalog No. |
022638581 |
| Maximum Speed |
12,000 rpm |
| Maximum RCF |
15,350 x g |
High-capacity Swing-bucket Rotor for Centrifuges 5810 and 5810 R Only - with Four 250 ml Buckets
- Adaptors accept up to 36 x 15 ml conical tubes
- Many different Adaptor types spin a variety of tubes per run
- Switch to microplate buckets for spinning deepwell plates and blocks (see below)
|

Swing-bucket Rotor,for standard tubes, includes 4 x 250 ml rectangular
buckets
Type A-4-62
|
Specifications |
|
With Four 250 ml Buckets |
| Catalog No. |
022638009 |
| Maximum Capacity |
4 x 250 ml |
| Maximum Speed (rpm) |
4,000 |
| Maximum RCF |
3,200 x g |
| Maximum Radius (mm) |
180 |
| Maximum Vessel Diameter (mm) |
62 |
| Maximum Vessel Length (mm) |
140 |
| Tube |
| |
 |
 |
 |
 |
 |
| Tube capacity (ml) |
3–5 |
1.5–2 |
4–7 |
3–15 |
7–17 |
| Typical tube size** |
10 x 75 |
11 x 38 |
12 x 75 or 13 x 100 |
16 x 100 |
17 x 100 |
| Max. tube length with aerosol-tight cover*** |
115 |
118 |
118 |
116 |
114 |
| Max. tube length without aerosol-tight cover*** |
123 |
122 |
121 |
121 |
118 |
| Adaptor for 250 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
 |
| Catalog No. |
022638203 |
022638220 |
022638246 |
022638262 |
022638301 |
| Number of tubes per Adaptor/per rotor |
25/100 |
16/64 |
15/60 |
12/48 |
12/48 |
| Adaptor bore (dia. mm) |
11 |
11 |
13 |
16 |
17.5 |
| Max. RCF (g-force) |
3,072 |
3,036 |
3,072 |
3,072 |
3,072 |
| Centrifugation radius (mm) |
173 |
171 |
173 |
173 |
173 |
| Tube |
| |
 |
|
 |
 |
**
|
| Tube capacity (ml) |
15 ml conical |
7–18 |
12–30 |
50 ml conical |
50 ml conical |
| Typical tube size** |
17 x 120 |
18 x 100 |
25 x 92 |
30 x 115 |
30 x 115 |
| Max. tube length with aerosol-tight cover*** |
121 |
119 |
116 |
116 |
— |
| Max. tube length without aerosol-tight cover*** |
127 |
129 |
119 |
122 |
122 |
| Adaptor for 250 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
 |
| Catalog No. |
022638289 |
022638327 |
022638360 |
022638343 |
022638351 |
| Number of tubes per Adaptor/per rotor |
9/36 |
8/32 |
4/16 |
3/12 |
4/16 |
| Adaptor bore (dia. mm) |
17.5 |
20 |
26 |
31 |
31 |
| Max. RCF (g-force) |
3,161 |
3,072 |
3,072 |
3,161 |
3,072 |
| Centrifugation radius (mm) |
178 |
173 |
173 |
178 |
173 |
| Tube |
| |
 |
 |
|
 |
|
| Tube capacity (ml) |
30–50 |
50–75 |
80–120 |
180–250 |
|
| Typical tube size** |
30 x 107 |
34 x 100 |
44 x 100 |
62 x 125 |
|
| Max. tube length with aerosol-tight cover*** |
113 |
118 |
125 |
127 |
|
| Max. tube length without aerosol-tight cover*** |
115 |
122 |
138 |
136 |
|
| Adaptor for 250 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
 |
Set of 2
|
Set of 2
 |
Set of 2
|
|
| Catalog No. |
022638386 |
022638408 |
022638424 |
022638441 |
|
| Number of tubes per Adaptor / per rotor |
4 / 16 |
2 / 8 |
1 / 4 |
1 / 4 |
|
| Adaptor bore (dia. mm) |
31 |
35 |
45 |
62 |
|
| Max. RCF (g-force) |
3,072 |
3,072 |
3,072 |
3,200 |
|
| Centrifugation radius (mm) |
173 |
173 |
173 |
180 |
|
Max. rotational speed: 4,000 rpm Max. rotational speed for 4°C (5810 R): 4,000 rpm
Temperature at max. rotational speed (5810 R): <4°C****
| * |
Swing-bucket rotors are delivered with rectangular buckets of one weight category. When placing follow-up orders for rectangular buckets, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In a rectangular bucket, the overall weight of the Adaptors and test tubes used must not exceed 630 g. |
| ** |
Caution: When 50 ml Falcon® tubes are used in Adaptor 22 63 835-1, the aerosol-tight cap can no longer be attached. |
| *** |
Diameter x length in mm. |
| **** |
Exceeding the maximum tube length may reduce the loading capacity of the Adaptor. Please carry out a manual swing test. |
| ***** |
23°C room temperature and 1 hour run time. |
| |
Adaptors, aerosol-tight covers, and rotors are autoclavable: 20 min. at 121°C. |
With Four Microplate Buckets
- Spins up to 16 microplates per run for high throughput

Swing-bucket Rotor, includes
4 microplate buckets
Type A-4-62-MTP
|
Specifications |
| |
With Four Microplate Buckets |
| Catalog No. |
022638041 |
| Maximum Capacity |
16 Microplate or 4 Deepwell Blocks |
| Maximum Speed (rpm) |
4,000 |
| Maximum RCF |
2,750 x g |
| Maximum Radius (mm) |
154 |
| Maximum Height (mm) |
45 |
- Max. rotational speed: 4,000 rpm
- Max. rotational speed for 4°C (5810 R): 4,000 rpm
- Temperature at max. rotational speed (5810 R): <0°C**
| * |
Swing-bucket rotors are delivered with microplate buckets of one weight category. Whenplacing follow-up orders for a replacementmicroplate bucket (Catalog No. 22 63 807-6),please state the existing weight category, whichis printed on the inside of the groove (e.g. 86,with the last two figures in grams). In the caseof plates used in the microplate buckets ofSwing-bucket Rotor 22 63 800-9 and/or 22 63804-1 (commercially available microplates,culture plates, and deep-well plates), themaximum weight must not exceed 380 g and themaximum height must not exceed 45 mm.
If plates other than the above-mentionedplates are used, the user must consult Brinkmannto confirm the suitability of using theseplates.The Swing-bucket Rotor 22 63 800-9 can beloaded with rectangular buckets or microplatebuckets. If the Swing-bucket Rotor is alreadyavailable with rectangular buckets, fourmicroplate buckets may be ordered and insertedinto the same rotor cross, for user-friendlyexchange of rectangular buckets and microplatebuckets
|
| ** |
23°C room temperature and 1 hour run time. |
| |
Buckets and rotors are autoclavable 20 min. at 121°C. |
Super High-capacity Swing-bucket Rotor for Centrifuges 5810 and 5810 R
With four 400 ml buckets
- High capacity for standard tubes; modular Adaptors for bottles and tubes
- Accommodates up to 4 x 400 ml bottles, 48 x 15 ml conical tubes, or 100 x 7 ml tubesl Ideal for medical and biotechnology laboratories

Swing-bucket Rotor, for standard tubes, includes 4 x 400 ml rectangular buckets
Type A-4-81 |
Specifications |
| Catalog No. |
022638602 |
| Maximum Capacity |
4 x 400 ml |
| Maximum Speed |
4,000 rpm |
| Maximum RCF |
3,250 x g |
| Maximum Radius |
180 mm |
| Maximum Vessel Diameter |
81 mm |
| Maximum Vessel Length |
133 mm |
| Tube |
|
|
 |
 |
 |
 |
 |
| Tube capacity (ml) |
2.6–7 |
7–17 |
15 ml conical |
50 ml conical |
50 ml Centriprep®
or skirted conical
(not shown) |
| Max. tube length with aerosol-tight cover** |
107 |
118 |
119 |
116** |
— |
| Max. tube length without aerosol-tight cover** |
108 |
118 |
121 |
122 |
121 |
| Adaptor for Flex-buckets for use with IsoRacks and cell culture flasks |
| Adaptor |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
 |
| Catalog No. |
022638700 |
022638726 |
022638742 |
022638769 |
022638904 |
| Number
of tubes per Adaptor/per rotor |
25/100 |
16/64 |
12/48 |
5/20 |
5/20 |
| Adaptor bore (dia. mm) |
13 |
17.5 |
17.5 |
31 |
31 |
| Max.
RCF (g-force) |
3,000 |
3,000 |
3,100 |
3,100 |
3,100 |
| Centrifugation radius (mm) |
168 |
168 |
173 |
173 |
173 |
Max. rotational
speed (4?°C are maintained securely in the 5810
R at a room temperature of 23°C) |
4,000 |
4,000 |
4,000 |
4,000 |
4,000 |
| Bottle |
| |
 |
 |
|
|
|
| Tube capacity (ml) |
180-2502 |
400 ml bottle
|
|
|
|
| Max. tube length with aerosol-tight cover** |
— |
— |
|
|
|
| Max. tube length without aerosol-tight cover** |
133 |
133 |
|
|
|
| Adaptor for Flex-buckets for use with IsoRacks and cell culture flasks |
| |
Set of 2
 |
Set of 2
 |
|
|
|
| Catalog No. |
022638921 |
022638785 |
|
|
|
| Number of tubes per Adaptor/per rotor |
1/4 |
1/4 |
|
|
|
| Adaptor bore (dia. mm) |
62 |
81 |
|
|
|
| Max. RCF (g-force) |
3,100 |
3,250 |
|
|
|
| Centrifugation radius (mm) |
173 |
180 |
|
|
|
Max. rotational speed (4?°C are maintained securely in the 5810 R at a room temperature of 23°C) |
4,000 |
4,000 |
|
|
|
- Max. rotational speed: 4,000 rpm
- Max. rotational speed for 4°C (5810 R): 3,900 rpm
- Temperature at max. rotational speed (5810 R): <8°C***
| * |
Swing-bucket rotors are delivered with buckets of one weight category. When placing follow-up orders for buckets, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In a bucket, the overall weight of the Adaptors and test tubes used must not exceed 800 g. |
| ** |
Exceeding the maximum tube length may reduce the loading capacity of the Adaptor. Please carry out a manual swing test. |
| *** |
23°C room temperature and 1 hour run time. |
| |
|
| |
Adaptors, aerosol-tight covers, and rotors are autoclavable: 20 min. at 121°C. |
With Four Microplate Buckets
- Accommodates up to 16 microplates, 2 culture plates,
4 deepwell plates, or Perfectprep® Plasmid purification kit assemblies
- Swing-bucket Rotor with microplate buckets;max. rotational speed: 4,000 rpm (max. RCF: 2,900 x g)

Swing-bucket Rotor, includes
4 microplate flex buckets
Type A-4-81-MTP |
Specifications |
| Catalog No. |
022638807 |
| Maximum Capacity |
2 Deepwell Assemblies (filter and catch-plate stacks) |
| Maximum Speed |
4,000 rpm |
| Maximum RCF |
2,900 x g |
| Type |
Microplate |
Cell culture |
DWP |
Kits |
| |
 |
 |
 |
 |
| Number per bucket / per rotor |
4 / 16 |
2 / 8 |
1 / 4 |
1 / 4 |
| Max. RCF (g-force) |
2,900 |
2,900 |
2,900 |
2,900 |
| Max. Centrifugation radius (mm) |
163 |
163 |
163 |
163 |
Max. rotational speed
(4°C are maintained securely in the 5810 R at a room temperature of 23°C) |
4,000 |
4,000 |
4,000 |
4,000 |
| Dimensions: Length x Width x Height (mm) |
128 x 85.6 x 15 |
128 x 85.6 x 21 |
128 x 85.6 x 45 |
128 x 85.6 x 60 |
- Max. rotational speed: 4,000 rpm
- Max. rotational speed for 4°C (5810 R): 4,000 rpm
- Temperature at max. rotational speed (5810 R): <0°C (+23°C room temperature and 1 hour run time)
|
| Adaptor for Flex-buckets for use with IsoRacks and cell culture flasks |
| |
 |
 |
|
|
| Tubes per Adaptor/per rotor |
24 / 96 |
24 / 96 |
|
|
| Max. RCF (g-force) |
2,700 |
2,600 |
|
|
| Max. Centrifugation radius (mm) |
150 |
146 |
|
|
Max. rotational speed
(4°C are maintained securely in the 5810 R at a room temperature of 23°C) |
4,000 |
4,000 |
|
|
- To use Rotors A-4-81/A-4-81-MTP with Centrifuge 5810/5810 R that were built prior to 2002, a minor electrical modification may be necessary.
- In the A-4-81-MTP bucket, plates should not exceed a weight of 380 g and a total height of 53 mm or 60 mm, respectively
|
|
 |