Discover More: Lomb Scientific

 
  • Eppendorf Pipettes +
           All  Liquid  Handling
  • Eppendorf Support Center
  • Centrifugation
  • Service & Repairs
  • PCR
  • Sample Preparation
  • Consumables
  • UV/VIS Detection
  • Cell Technology


  • Visit Lomb Scientifics' dedicated Eppendorf Section


  • Polystainer - IUL
  •  

    Eppendorf Multipurpose Centrifuge Model 5804


    description specifications rotors


    Rotors and Adaptors for Multipurpose Centrifuges 5804, 5804 R, 5810, and 5810 R

    Eppendorf Multipurpose Centrifuges accommodate fixed-angle and swing-bucket rotors for a variety of applications. A large selection of Adaptors allows maximum application flexibility with a minimum number of rotors.

    Product features
    • All rotors, buckets, and Adaptors are autoclavable at 121°C
    • Stackable modular Adaptors for swing-bucket rotors are easy to disassemble for vessels of different lengths, for cleaning and autoclavin
    • g
    • Adaptors are color coded for quick and simple identification
    • Buckets can be easily sealed with aerosol-tight covers to provide protection against biohazardous samples

    Quick Find links:


    High-Speed Fixed-angle 30-place Microcentrifuge Rotors
    • Spin up to 20,800 x g
    • Quick release lid

    Aerosol-tight Fixed-angle Rotor,
    for microcentrifuge tubes
    Type FA-45-30-11

    Fixed-angle Rotor,
    for microcentrifuge tubes
    Type F-45-30-11

    Specifications
    Catalog No.
    – Standard
    – Aerosol-tight
    022637002
    022637100
    Maximum Capacity 30 x 2.0 ml
    Maximum Speed 14,000 rpm
    Maximum RCF 20,800 x g
    Maximum Radius 95 mm
    Maximum Vessel Diameter 11.5 mm
    Maximum Vessel Length 50 mm

    Tube

    (Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.)

    Tube capacity (ml) 1.5–2 0.2 0.4 0.5
    Adaptor for 30 x 1.5 ml/2.0 ml fixed-angle rotor for use with PCR and micro test tubes
    Adaptor Can be inserted without Adaptor Set of 6
    Set of 6
    Set of 6
    Catalog No. 022636260 022636243 022636227
    Number of tubes per Adaptor/per rotor «/30 1/30 1/30 1/30
    Adaptor bore (dia. mm) 6 6 8
    Max. RCF (g-force) 20,800 16,200 20,800 18,400
    Centrifugation radius (mm) 95 74 95 84
    • Max. rotational speed: 14,000 rpm
    • Max. rotational speed for 4°C (5804 R/5810 R): 14,000 rpm
    • Temperature at max. rotational speed (5804 R, 230 V/5810 R, 230 V): <4°C*
    * 23°C room temperature and 1 hour run time.
    Adaptors, aerosol-tight covers, and rotors are autoclavable: 20 min. at 121°C.

        return to Quick Find links list    


    High-speed Fixed-angle Rotor for 6 x 85 ml Tubes
    • Adaptors accommodate 1.5 to 50 ml tubes
    • Spins conical tubes at high speeds

    High-speed Fixed-angle Rotor,
    6 x 85 ml
    Type F-34-6-38
    Specifications
    Catalog No. 022637207
    Maximum Capacity 6 x 85 ml
    Maximum Speed 11,000 rpm*
    Maximum RCF 15,500 x g*
    Maximum Radius 115 mm
    Maximum Vessel Diameter 38 mm
    Maximum Vessel Length 120 mm
    *12,000 rpm/18,500 x g in Model 5810 R. To maintain 4°C with this rotor, a maximum speed of 10,500 rpm is required.

    Tube
    (Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.)
    Tube capacity (ml) 1.5–2 7–15 15 ml conical 15–18
    Max. tube length for all positions below the rotor lid (mm)* 110 112 123 123
    Adaptor for 6 ml x 85 ml fixed-angle rotor for use with Standard, conical and micro test tubes
    Adaptor Set of 2
    Set of 2
    Set of 2
    Set of 2

    Catalog No. 022637215 022637223 022637274 022637231
    Number of tubes per Adaptor/per rotor 4/24 2/12 1/6 1/6
    Adaptor bore (dia. mm) 11 16 17 18
    Max. RCF (g-force) at 11,000 rpm 15,290 15,150 14,470 14,750
    Max. RCF (g-force) at 12,000 rpm – 5810 R 18,190 18,030 17,230 17,550
    Centrifugation radius (mm) 113 112 107 109
    Tube
    (Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.)
    Tube capacity (ml) 20–30 50 50 ml conical 85
    Max. tube length for all positions below the rotor lid (mm)* 123 123 121 121
    Adaptor for 6 ml x 85 ml fixed-angle rotor for use with Standard, conical and micro test tubes
    Adaptor Set of 2
    Set of 2
    Set of 2
    Can be inserted without Adaptor
    Catalog No. 022637240 022637258 022637266
    Number of tubes per Adaptor/per rotor 1/6 1/6 1/6 –/6
    Adaptor bore (dia. mm) 26 29 29.5 38
    Max. RCF (g-force) at 11,000 rpm 14,880 15,150 14,610 15,560
    Max. RCF (g-force) at 12,000 rpm – 5810 R 17,710 18,030 17,390 18,500
    Centrifugation radius (mm) 110 112 108 115

    • Temperature at max. rotational speed (12,000 rpm), 5810 R (230 V): <10°C**
    • Temperature at max. rotational speed (12,000 rpm), 5810 R (120 V): <12°C**
    • Max. rotational speed for 4°C, 5810 R: 10,500 rpm
    • Temperature at max. rotational speed (11,000 rpm), 5804 R: <6°C**
    • Max. rotational speed for 4°C, 5804 R: 10,500 rpm

    * Exceeding the maximum tube length may reduce the loading capacity of the Adaptor.
    ** 23°C room temperature and 1 hour run time.
      Adaptors, aerosol-tight covers, and rotors are autoclavable 20 min. at 121°C.

        return to Quick Find links list    


    Horizontal Drum Rotor
    • Accepts up to 60 x 1.5 ml or 120 x 0.4 ml tubes for high throughput
    • Carriers can be used as tube racks for easy handling

    Drum Rotor,
    Type T-60-11
    Specifications
    Catalog No. 022638505
    Maximum Capacity 60 x 1.5–2.0 ml, 120 x 0.4 ml
    Maximum Speed 14,000 rpm
    Maximum RCF 16,435 x g
    Maximum Radius 750 mm
    Maximum Vessel Diameter 11 mm
    Maximum Vessel Length 41 mm

        return to Quick Find links list    


    PCR/MTP/DWP Swing-bucket Rotor with Two Buckets
    • Ideal for deepwell blocks and purification kit assemblies with heights greater than 45mm
    • Use with Eppendorf Perfectprep® Plasmid 96 Spin Kits
    • Spins up to 10 microplates or 4 deepwell plates

    Deepwell plate Rotor, includes
    2 Deepwell plate buckets
    Type A-2-DWP
    Specifications
    Catalog No. 022638564
    Maximum Capacity 2 Deepwell Assemblies
    (filter and catch-plate stacks)
    Maximum Speed 3,700 rpm
    Maximum RCF 2,250 x g
    Maximum Radius 147 mm
    Maximum Vessel Dimensions, W x D 91 x 134 mm
    Maximum Height 85 mm

    Type Microplate Cell culture DWP Kits
    Number per bucket/per rotor 5/10 4/8 2/4 1/4
    Max. RCF (g-force) 2,250 2,250 2,250 2,250
    Max. Centrifugation radius (mm) 147 147 147 147
    Dimensions: Length x Width x Height (mm) 128 x 85.6 x 15 128 x 85.6 x 21 128 x 85.6 x 45 128 x 85.6 x 60
    • Max. rotational speed: 3,700 rpm
    • Max. rotational speed for 4°C (5804 R / 5810 R): 3,700 rpm
    • Temperature at max. rotational speed (5804 R / 5810 R): <0°C**
    *  

    The deepwell plate rotor 22 63 856-4 is delivered with buckets of one weight category. When placing follow-up orders for a deepwell plate bucket, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In the case of plates used in deepwell plate buckets for rotor 22 63 856-4 (commercially available microplates, culture plates, and deepwell plates), the maximum weight must not exceed 380 g and the maximum height must not exceed 85 mm. If plates other than the above-mentioned plates are used, the user must consult Eppendorf to confirm the suitability of using these plates.

    **   23°C room temperature and 1 hour run time.
       
      Buckets and rotors are autoclavable 20 min. at 121°C.

        return to Quick Find links list    


    Swing-bucket Rotor with Four 100 ml Buckets
    • Adaptors accommodate 0.25 to 100 ml tubes
    • Accepts four 2 x 50 ml buckets to spin eight 50 ml conicals>
    • Spins up to 4,500 x g for quick separation

    Swing-bucket Rotor,for
    standard tubes, includes 4 x 100 ml rectangular buckets
    Type A-4-44
    Specifications
    Catalog No. 022637401
    Maximum Capacity 4 x 100 ml
    Maximum Speed 5,000 rpm
    Maximum RCF 4,500 x g
    Maximum Radius 161 mm
    Maximum Vessel Diameter 46 mm
    Maximum Vessel Length 125 mm

    Tube

     

    Tube capacity (ml) 3–5 1.5–2 4–7 3–15 7–17
    Typical tube size** 10 x 75 11 x 38 12 x 75 or 13 x 100 16 x 100 17 x 100
    Max. tube length with aerosol-tight cover*** 102 105 106 106 106
    Max. tube length without aerosol-tight cover*** 105 111 108 108 110
    Adaptor for 100 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes
    Adaptor Set of 2

    Set of 2

    Set of 2

    Set of 2
    Set of 2
    Catalog No. 022637509 022637525 022637541 022637568 022637584
    Number of tubes per Adaptor/per rotor 14/56 12/48 9/36 7/28 6/24
    Adaptor bore (dia. mm) 11 11 13 16 17.5
    Max. RCF (g-force) 4,192 4,136 4,192 4,192 4,192
    Centrifugation radius (mm) 150 148 150 150 150
    Tube
    For use with aerosol-tight cover
    Tube capacity (ml) 15 ml conical 15 ml conical 7–18 12–30 50 ml conical
    Typical tube size** 17 x 120 17 x 120 18 x 100 25 x 92 30 x 115
    Max. tube length with aerosol-tight cover*** 109 121 104 100 112
    Max. tube length without aerosol-tight cover*** 121 121 107 108 122
    Adaptor for 100 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes
    Adaptor Set of 2
    Set of 2
    Set of 2
    Set of 2
    Set of 2
    Catalog No. 022637606 022637614 022637622 022637649 022637665
    Number of tubes per Adaptor/per rotor 4/16 2/8 4/16 2/8 1/4
    Adaptor bore (dia. mm) 17.5 17.5 20 26 31
    Max. RCF (g-force) 4,332 4,388 4,192 4,192 4,332
    Centrifugation radius (mm) 155 157 150 150 155
    Tube
    For use with aerosol-tight cover
    Tube capacity (ml) 50 ml conical 50 ml conical 30–50 50–75 80–120
    Typical tube size** 30 x 115 30 x 115 30 x 107 34 x 100 44 x 100
    Max. tube length with aerosol-tight cover*** 119 108 108 100
    Max. tube length without aerosol-tight cover*** 119 119 122 119 114
    Adaptor for 100 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes
    Adaptor Set of 2
    Bucket
    Set of 2
    Set of 2
    Set of 2
    Catalog No. 022637673 022637452 022637681 022637703 022637720
    Number of tubes per Adaptor/per rotor 1/4 2/8 1/4 1/4 1/4
    Adaptor bore (dia. mm) 31 31 35 45
    Max. RCF (g-force) 4,388 4,500 4,192 4,192 4,192
    Centrifugation radius (mm) 157 161 150 150 150
    • Max. rotational speed: 5,000 rpm
    • Max. rotational speed for 4°C (5804 R / 5810 R): 4,800 rpm
    • Temperature at max. rotational speed (5804 R / 5810 R): 6°C****
    *  

    Swing-bucket rotors are delivered with rectangular buckets of one weight category. When placing follow-up orders for rectangular buckets, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In a rectangular bucket, the overall weight of the Adaptors and test tubes used must not exceed 310 g.

    **   Diameter x length in mm.
    ***   Exceeding the maximum tube length may reduce the loading capacity of the Adaptor. Please carry out a manual swing test.
    ****   23°C room temperature and 1 hour run time.
      Adaptors, aerosol-tight covers, and rotors are autoclavable: 20 min. at 121°C.

        return to Quick Find links list    

    PCR Strip Rotor
    • Max. rotational speed 10,000 rpm
    • Up to 6 x 8-tube strips, 6 x 5-tube strips or 48 x 0.2 ml PCR single tubes
    • Safe and easy handling without additional lid

    Type F-45-48-PCR
    Specifications
    Catalog No. 022638581
    Maximum Speed 12,000 rpm
    Maximum RCF 15,350 x g

        return to Quick Find links list