Eppendorf Multipurpose Centrifuge Model 5804 |
|
Rotors and Adaptors for Multipurpose Centrifuges 5804, 5804 R, 5810, and 5810 R
Eppendorf Multipurpose Centrifuges accommodate fixed-angle and swing-bucket rotors for a variety of applications. A large selection of Adaptors allows maximum application flexibility with a minimum number of rotors.
Product features
- All rotors, buckets, and Adaptors are autoclavable at 121°C
- Stackable modular Adaptors for swing-bucket rotors are easy to disassemble for vessels of different lengths, for cleaning and autoclavin
g
- Adaptors are color coded for quick and simple identification
- Buckets can be easily sealed with aerosol-tight covers to provide protection against biohazardous samples
High-Speed Fixed-angle 30-place Microcentrifuge Rotors
- Spin up to 20,800 x g
- Quick release lid

Aerosol-tight Fixed-angle Rotor, for microcentrifuge tubes
Type FA-45-30-11

Fixed-angle Rotor,
for microcentrifuge tubes Type F-45-30-11 |
Specifications |
Catalog No. – Standard – Aerosol-tight |
022637002
022637100 |
| Maximum Capacity |
30 x 2.0 ml |
| Maximum Speed |
14,000 rpm |
| Maximum RCF |
20,800 x g |
| Maximum Radius |
95 mm |
| Maximum Vessel Diameter |
11.5 mm |
| Maximum Vessel Length |
50 mm |
- Max. rotational speed: 14,000 rpm
- Max. rotational speed for 4°C (5804 R/5810 R): 14,000 rpm
- Temperature at max. rotational speed (5804 R, 230 V/5810 R, 230 V): <4°C*
| * |
23°C room temperature and 1 hour run time.
|
 |
Adaptors, aerosol-tight covers, and rotors are autoclavable: 20 min. at 121°C. |
High-speed Fixed-angle Rotor for 6 x 85 ml Tubes
- Adaptors accommodate 1.5 to 50 ml tubes
- Spins conical tubes at high speeds

High-speed Fixed-angle Rotor, 6 x 85 ml
Type F-34-6-38 |
Specifications |
| Catalog No. |
022637207 |
| Maximum Capacity |
6 x 85 ml |
| Maximum Speed |
11,000 rpm* |
| Maximum RCF |
15,500 x g* |
| Maximum Radius |
115 mm |
| Maximum Vessel Diameter |
38 mm |
| Maximum Vessel Length |
120 mm |
| *12,000 rpm/18,500 x g in Model 5810 R. To maintain 4°C with this rotor, a maximum speed of 10,500 rpm is required. |
| Tube |
|
(Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.) |
 |
 |
 |
 |
| Tube capacity (ml) |
1.5–2 |
7–15 |
15 ml conical |
15–18 |
| Max. tube length for all positions below the rotor lid (mm)* |
110 |
112 |
123 |
123 |
| Adaptor for 6 ml x 85 ml fixed-angle rotor for use with Standard, conical and micro test tubes |
| Adaptor |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Set of 2
|
| Catalog No. |
022637215 |
022637223 |
022637274 |
022637231 |
| Number
of tubes per Adaptor/per rotor |
4/24 |
2/12 |
1/6 |
1/6 |
| Adaptor bore (dia. mm) |
11 |
16 |
17 |
18 |
| Max.
RCF (g-force) at 11,000 rpm |
15,290 |
15,150 |
14,470 |
14,750 |
| Max.
RCF (g-force) at 12,000 rpm – 5810 R |
18,190 |
18,030 |
17,230 |
17,550 |
| Centrifugation radius (mm) |
113 |
112 |
107 |
109 |
| Tube |
| (Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.) |
 |
 |
 |
 |
| Tube capacity (ml) |
20–30 |
50 |
50 ml conical |
85 |
| Max. tube length for all positions below the rotor lid (mm)* |
123 |
123 |
121 |
121 |
| Adaptor for 6 ml x 85 ml fixed-angle rotor for use with Standard, conical and micro test tubes |
| Adaptor |
Set of 2
 |
Set of 2
 |
Set of 2
 |
Can be inserted without Adaptor |
| Catalog No. |
022637240 |
022637258 |
022637266 |
— |
| Number of tubes per Adaptor/per rotor |
1/6 |
1/6 |
1/6 |
–/6 |
| Adaptor bore (dia. mm) |
26 |
29 |
29.5 |
38 |
| Max. RCF (g-force) at 11,000 rpm |
14,880 |
15,150 |
14,610 |
15,560 |
| Max. RCF (g-force) at 12,000 rpm – 5810 R |
17,710 |
18,030 |
17,390 |
18,500 |
| Centrifugation radius (mm) |
110 |
112 |
108 |
115 |
- Temperature at max. rotational speed (12,000 rpm), 5810 R (230 V): <10°C**
- Temperature at max. rotational speed (12,000 rpm), 5810 R (120 V): <12°C**
- Max. rotational speed for 4°C, 5810 R: 10,500 rpm
- Temperature at max. rotational speed (11,000 rpm), 5804 R: <6°C**
- Max. rotational speed for 4°C, 5804 R: 10,500 rpm
| * |
Exceeding the maximum tube length may reduce the loading capacity of the Adaptor. |
| ** |
23°C room temperature and 1 hour run time. |
 |
 |
| |
Adaptors, aerosol-tight covers, and rotors are autoclavable 20 min. at 121°C. |
Horizontal Drum Rotor
- Accepts up to 60 x 1.5 ml or 120 x 0.4 ml tubes for high throughput
- Carriers can be used as tube racks for easy handling

Drum Rotor,
Type T-60-11 |
Specifications |
| Catalog No. |
022638505 |
| Maximum Capacity |
60 x 1.5–2.0 ml, 120 x 0.4 ml |
| Maximum Speed |
14,000 rpm |
| Maximum RCF |
16,435 x g |
| Maximum Radius |
750 mm |
| Maximum Vessel Diameter |
11 mm |
| Maximum Vessel Length |
41 mm |
PCR/MTP/DWP Swing-bucket Rotor with Two Buckets
- Ideal for deepwell blocks and purification kit assemblies with heights greater than 45mm
- Use with Eppendorf Perfectprep® Plasmid 96 Spin Kits
- Spins up to 10 microplates or 4 deepwell plates

Deepwell plate Rotor, includes
2 Deepwell plate buckets
Type A-2-DWP |
Specifications |
| Catalog No. |
022638564 |
| Maximum Capacity |
2 Deepwell Assemblies (filter and catch-plate stacks) |
| Maximum Speed |
3,700 rpm |
| Maximum RCF |
2,250 x g |
| Maximum Radius |
147 mm |
| Maximum Vessel Dimensions, W x D |
91 x 134 mm |
| Maximum Height |
85 mm |
| Type |
Microplate |
Cell culture |
DWP |
Kits |
 |
 |
 |
 |
 |
| Number per bucket/per rotor |
5/10 |
4/8 |
2/4 |
1/4 |
| Max. RCF (g-force) |
2,250 |
2,250 |
2,250 |
2,250 |
| Max. Centrifugation radius (mm) |
147 |
147 |
147 |
147 |
| Dimensions: Length x Width x Height (mm) |
128 x 85.6 x 15 |
128 x 85.6 x 21 |
128 x 85.6 x 45 |
128 x 85.6 x 60 |
- Max. rotational speed: 3,700 rpm
- Max. rotational speed for 4°C (5804 R / 5810 R): 3,700 rpm
- Temperature at max. rotational speed (5804 R / 5810 R): <0°C**
| * |
The deepwell plate rotor 22 63 856-4 is delivered with buckets of one weight category. When placing follow-up orders for a deepwell plate bucket, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In the case of plates used in deepwell plate buckets for rotor 22 63 856-4 (commercially available microplates, culture plates, and deepwell plates), the maximum weight must not exceed 380 g and the maximum height must not exceed 85 mm. If plates other than the above-mentioned plates are used, the user must consult Eppendorf to confirm the suitability of using these plates. |
| ** |
23°C room temperature and 1 hour run time. |
| |
|
| |
Buckets and rotors are autoclavable 20 min. at 121°C. |
|
Swing-bucket Rotor with Four 100 ml Buckets
- Adaptors accommodate 0.25 to 100 ml tubes
- Accepts four 2 x 50 ml buckets to spin eight 50 ml conicals>
- Spins up to 4,500 x g for quick separation

Swing-bucket Rotor,for
standard tubes, includes 4 x 100 ml rectangular buckets
Type A-4-44 |
Specifications |
| Catalog No. |
022637401 |
| Maximum Capacity |
4 x 100 ml |
| Maximum Speed |
5,000 rpm |
| Maximum RCF |
4,500 x g |
| Maximum Radius |
161 mm |
| Maximum Vessel Diameter |
46 mm |
| Maximum Vessel Length |
125 mm |
| Tube |
|
|
 |
 |
 |
 |
 |
| Tube capacity (ml) |
3–5 |
1.5–2 |
4–7 |
3–15 |
7–17 |
| Typical tube size** |
10 x 75 |
11 x 38 |
12 x 75 or 13 x 100 |
16 x 100 |
17 x 100 |
| Max. tube length with aerosol-tight cover*** |
102 |
105 |
106 |
106 |
106 |
| Max. tube length without aerosol-tight cover*** |
105 |
111 |
108 |
108 |
110 |
| Adaptor for 100 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
|
Set of 2
|
Set of 2
|
Set of 2
|
Set of 2
 |
| Catalog No. |
022637509 |
022637525 |
022637541 |
022637568 |
022637584 |
| Number of tubes per Adaptor/per rotor |
14/56 |
12/48 |
9/36 |
7/28 |
6/24 |
| Adaptor bore (dia. mm) |
11 |
11 |
13 |
16 |
17.5 |
| Max. RCF (g-force) |
4,192 |
4,136 |
4,192 |
4,192 |
4,192 |
| Centrifugation radius (mm) |
150 |
148 |
150 |
150 |
150 |
| Tube |
 |
 |
For use with aerosol-tight cover
 |
|
 |
 |
| Tube capacity (ml) |
15 ml conical |
15 ml conical |
7–18 |
12–30 |
50 ml conical |
| Typical tube size** |
17 x 120 |
17 x 120 |
18 x 100 |
25 x 92 |
30 x 115 |
| Max. tube length with aerosol-tight cover*** |
109 |
121 |
104 |
100 |
112 |
| Max. tube length without aerosol-tight cover*** |
121 |
121 |
107 |
108 |
122 |
| Adaptor for 100 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
|
Set of 2
|
Set of 2
 |
Set of 2
|
Set of 2
|
| Catalog No. |
022637606 |
022637614 |
022637622 |
022637649 |
022637665 |
| Number of tubes per Adaptor/per rotor |
4/16 |
2/8 |
4/16 |
2/8 |
1/4 |
| Adaptor bore (dia. mm) |
17.5 |
17.5 |
20 |
26 |
31 |
| Max. RCF (g-force) |
4,332 |
4,388 |
4,192 |
4,192 |
4,332 |
| Centrifugation radius (mm) |
155 |
157 |
150 |
150 |
155 |
| Tube |
 |
For use with aerosol-tight cover
|
 |
 |
 |
|
| Tube capacity (ml) |
50 ml conical |
50 ml conical |
30–50 |
50–75 |
80–120 |
| Typical tube size** |
30 x 115 |
30 x 115 |
30 x 107 |
34 x 100 |
44 x 100 |
| Max. tube length with aerosol-tight cover*** |
119 |
— |
108 |
108 |
100 |
| Max. tube length without aerosol-tight cover*** |
119 |
119 |
122 |
119 |
114 |
| Adaptor for 100 ml rectangular bucket for use with Standard, Vacutainer®, conical tubes and micro test tubes |
| Adaptor |
Set of 2
 |
Bucket
|
Set of 2
|
Set of 2
|
Set of 2
 |
| Catalog No. |
022637673 |
022637452 |
022637681 |
022637703 |
022637720 |
| Number of tubes per Adaptor/per rotor |
1/4 |
2/8 |
1/4 |
1/4 |
1/4 |
| Adaptor bore (dia. mm) |
31 |
— |
31 |
35 |
45 |
| Max. RCF (g-force) |
4,388 |
4,500 |
4,192 |
4,192 |
4,192 |
| Centrifugation radius (mm) |
157 |
161 |
150 |
150 |
150 |
- Max. rotational speed: 5,000 rpm
- Max. rotational speed for 4°C (5804 R / 5810 R): 4,800 rpm
- Temperature at max. rotational speed (5804 R / 5810 R): 6°C****
| * |
Swing-bucket rotors are delivered with rectangular buckets of one weight category. When placing follow-up orders for rectangular buckets, please state the existing weight category, which is printed on the inside of the groove (e.g. 86, with the last two figures in grams). In a rectangular bucket, the overall weight of the Adaptors and test tubes used must not exceed 310 g. |
| ** |
Diameter x length in mm. |
| *** |
Exceeding the maximum tube length may reduce the loading capacity of the Adaptor. Please carry out a manual swing test. |
| **** |
23°C room temperature and 1 hour run time.
|
| |
Adaptors, aerosol-tight covers, and rotors are autoclavable: 20 min. at 121°C. |
PCR Strip Rotor
- Max. rotational speed 10,000 rpm
- Up to 6 x 8-tube strips, 6 x 5-tube strips or 48 x 0.2 ml PCR single tubes
- Safe and easy handling without additional lid
 Type F-45-48-PCR |
Specifications |
| Catalog No. |
022638581 |
| Maximum Speed |
12,000 rpm |
| Maximum RCF |
15,350 x g |
|
 |
|