Discover More: Lomb Scientific

 
  • Eppendorf Pipettes +
           All  Liquid  Handling
  • Eppendorf Support Center
  • Centrifugation
  • Service & Repairs
  • PCR
  • Sample Preparation
  • Consumables
  • UV/VIS Detection
  • Cell Technology


  • Visit Lomb Scientifics' dedicated Eppendorf Section


  • Polystainer - IUL
  •  

    Eppendorf Centrifuge Model 5702


    description specifications rotors

      Swing-bucket and Fixed-angle Rotors for Eppendorf Centrifuge 5702  

      Product features of all rotors  
    • Minimal heating of samples
    • Quick acceleration in 25 seconds or less
    • All rotors and accessories are autoclavable (121°C, 20 min.)
      A-4-38 Swing-bucket Rotor  
    • Choice of round buckets with optional aerosol-tight lids or high-capacity square buckets for round-bottom tubes
    • Round buckets accommodate conical and round-bottom tubes from 1.1 ml to 85 ml
    • Square buckets accommodate round- bottom tubes from 5 ml to 25 ml
    • Optional aerosol-tight covers protect the user from hazardous samples*

    *Aerosol containment has been tested and approved in accordance with the procedure of Bennett et al. from the Centre for Applied Microbiology & Research in Porton Down, CAMR, Great Britain, and according to annex AA of the IEC 1010-2-020 safety standard. The test procedure simulates the accidental release of biological material (bacteria) inside the closed rotor or closed rectangular bucket. Air and swab samples are taken to ensure that no material is released from the closed rectangular buckets or closed rotor during operation.

      F-35-30-17 Fixed-angle Rotor  
    • Available with a choice of 10 or 30 steel tube holders
    • Accommodates up to 20 x 15 ml conical tubes or 30 x 15 ml round-bottom tubes
      A-8-17 Swing-bucket Rotor  
    • Affordable and simple rotor design
    • Accommodates up to 8 x 15 ml conical or round-bottom tubes


    Specifications    
      With round buckets With square buckets
    Maximum Capacity 4 x 85 ml 8 x 25 ml
    Maximum Speed 4,400 rpm 4,400 rpm
    Maximum RCF 3,000 x g 2,750 x g
    Maximum Radius 137 mm 127 mm
    Maximum Vessel Diameter 38 mm 25 mm
    Maximum Vessel Length 121 mm 100 mm

    Swing-bucket Rotor includes 4 x 85 ml round buckets
    Type A-4-38
    022639048

    Round bucket(1) set of 4
    022639081

    Aerosol-tight cover, set of 2

    022639293

    Tube

    (Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.)

    1.1–1.4 2–7 2.6–7 4–10 9–15
    Max. tube length for all positions below the aerosol-tight lid (mm)* 100 100 100 100 100
    Max. tube length for all positions without aerosol-tight lid (mm)* 100 100 100 100 100
    Adapter Set of 2
    Set of 2
    Set of 2
    Set of 2
    Set of 2
    Catalog No. 022639285 022639102 022639242 022639269 022639129
    5/20 5/20 4/16 4/16 4/16
    Adapter bore (dia. mm) 8.5 12.5 13.5 16 17.5
    2,850 2,850 2,850 2,850 2,850
    Centrifugation radius (mm) 132 132 132 132 130
    4,400 4,400 4,400 4,400 4,400

    Tube

    (Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.)

    Tube capacity (ml) 15 ml conical 15 ml conical 25 50 ml conical 85
    Max. tube length for all positions below the aerosol-tight lid (mm)* 120 – 100 115 106
    Max. tube length for all positions without aerosol-tight lid (mm)* 121 121 100 115 106
    height=65>Adapter Set of 2
    Set of 2
    Set of 2
    Set of 2
    Set of 2
    Catalog No. 022639188 022639200(2) 022639145 022639226 022639161
    Number of tubes per adapter/per rotor 1/4 2/8 1/4 1/4 1/4
    Adapter bore (dia. mm) 17.2 17.2 25 30 38
    Max. RCF (g-force) 3,000 3,000 2,900 2,900 2,900
    Centrifugation radius (mm) 137 137 135 135 135
    Max. rotational speed (rpm) 4,400 4,400 4,400 4,400 4,400

    (1) Swing-bucket Rotors are supplied with buckets of one weight class. Adapters and tubes which together do not exceed a maximum weight of 190 g can be put in one round bucket. The total weight of a loaded bucket (including adapters and filled tubes) may not exceed 375 g. The weight given on the rotor applies to buckets including their load.

    (2) Cannot be use with aerosol-tight lids.

    *If maximum tube length is exceeded, the loading capacity of the adapter may be reduced. Please perform a manual swing test.

    Adapter, caps and rotor are autoclavable: 121°C, 20 min.

     
    Swing-bucket Rotor without buckets
    Type A-4-38
    022639064

    Rectangular bucket
    set of 4
    022639307

    Tube

    (Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.)

    Tube capacity (ml) 5–7 9 15 20 25
    Max. tube length for all positions (mm)* 100 100 100 100 100
    Adapter Set of 2
    Set of 2
    Set of 2
    Set of 2
    Set of 2
    Catalog No. 022639323 022639340 022639366 022639382 022639391
    Number of tubes per adapter/per rotor 10/40 8/32 6/24 4/16 2/8
    Adapter bore (dia. mm) 13 14.5 17.5 22 25
    Max. RCF (g-force) 2,750 2,750 2,750 2,750 2,750
    bgColor=#0e2a8a height=19>Centrifugation radius (mm) 127 127 127 127 127
    Max. rotational speed (rpm) 4,400 4,400 4,400 4,400 4,400

    *If maximum tube length is exceeded, the loading capacity of the adapter may be reduced. Please perform a manual swing test.



    Fixed-angle Rotor includes
    30 steel sleeves
    Type F-35-30-17
    022639404


    Fixed-angle Rotor includes
    10 steel sleeves
    Type F-35-30-17
    022639421


    Steel sleeve, set of 10
    022639439

    Tube

    (Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.)

    Tube capacity (ml) 15 15 ml conical
    Max. tube length for all positions (mm)* 115(3) inner ring: 135; outer ring: 120
    Catalog No. 022639455 022639447
    Number of tubes per rotor for versions with 30 sleeves 30 20
    Sleeves (dia. mm) 18 18
    Max. RCF (g-force) 2,750 2,700
    Centrifugation radius (mm) 127 125
    Max. rotational speed (rpm) 4,400 4,400

    (3) 105 with a load of conical tubes and round-bottom tubes.

      24-place microcentrifuge Rotor F-45-24-11  
    • Max. rcf: 1,700 x g
    • Low-noise centrifugation without rotor lid
    • Autoclavable (121°C, 20 min)l Centrifugation of 1.5 ml and 2 ml Eppendorf TubesŽ
    • Adapter for:
      – 0.2 ml PCR tubes
      – 0.4 ml centrifuge tubes
      – 0.5 ml micro test tubes and 0.6 ml MicrotainerŽ tubes

      18-place Screw-cap/cryovial Rotor F-45-18-17-Cryo  
    • Max. rcf: 1,840 x g
    • Low-noise centrifugation without rotor lid
    • Rotor and adapter autoclavable (121°C, 20 min)
    • Centrifugation of 1 ml–2 ml CryotubesŽ (max. 13 mm, can be used with adapter) and
      screw-cap centrifugation tubes (max. 17 mm, can be used directly)


    Swing-bucket Rotor,
    8 x 15 ml
    Type A-8-17
    022639501

    Tube

    (Please observe the limiting data from the tube manufacturer regarding the maximum load- bearing capacity of the tubes.)

    Tube capacity (ml) 15 15 ml conical
    Max. tube length for all positions (mm)* 120 120
    Catalog No. for adapters 022639510 022639528
    Number of tubes per rotor 8 8
    Sleeves (dia. mm) 18 18
    Max. RCF (g-force) 2,750 2,700
    Centrifugation radius (mm) 128 128
    Max. rotational speed (rpm) 4,400 4,400

    *If maximum tube length is exceeded, the loading capacity of the adapter may be reduced.